DS DSEN SO 10328 2007-JAN-18 Prosthetcs - Structural testng of lower-lmb prostheses - Requrements and test methods
DS DSEN SO 10328 2007-JAN-18 Prosthetcs - Structural testng of lower-lmb prostheses - Requrements and test methods
ISO 10328 specifies procedures for static and cyclic strength tests of lower-limb prostheses where, with one exception, compound loadings are produced by the application of a single test force. The compound loads in the test sample relate to the peak values of the components of loading which normally occur at different instants during the stance phase of walking.
IMPORTANT - This International Standard is suitable for the assessment of the conformity of lower
limb prosthetic devices/structures with the strength requirements specified in 4.4 of ISO 22523:2006
(see NOTE 1). Prosthetic ankle-foot devices and foot units on the market, which have demonstrated
their compliance with the strength requirements specified in 4.4 of EN 12523:1999 through submission
to the relevant tests of ISO 10328:1996, need not be retested to ISO 22675:2006.
WARNING - This International Standard is not suitable to serve as a guide for the selection of a
specific lower limb prosthetic device/structure in the prescription of an individual lower limb
prosthesis! Any disregard of this warning can result in a safety risk for amputees.
This International Standard specifies procedures for static and cyclic strength tests on lower-limb prostheses
(see NOTE 2) which typically produce compound loadings by the application of a single test force. The compound loads in the test sample relate to the peak values of the components of loading which normally occur at different instants during the stance phase of walking.
The tests described in this International Standard comprise
. principal static and cyclic tests for all components;
. a separate static test in torsion for all components;
. separate static and cyclic tests on ankle-foot devices and foot units for all ankle-foot devices as single components including ankle units or ankle attachments and all foot units as single components;
. a separate static ultimate strength test in maximum knee flexion on knee joints and associated parts for all knee units or knee-shin-assemblies and adjacent components that normally provide the flexion stop on a complete prosthesis;
. separate static and cyclic tests on knee locks for all mechanisms which lock the knee joint in the
extended position of the knee unit or knee-shin-assembly.
The tests described in this International Standard apply to specific types of ankle-disarticulation prostheses (see NOTE 3), to transtibial (below-knee), knee-disarticulation and transfemoral (above-knee) prostheses and to the distal (lower) part of hip-disarticulation and hemi-pelvectomy prostheses (see NOTE 4).
NOTE 1 ISO 22523 (formerly EN 12523) addresses those of the Essential Requirements listed in Annex 1 of the European Medical Device Directive 93/42/EEC that are applicable to external limb prostheses and external orthoses.
NOTE 2 The tests can be performed on complete structures, on part structures or on individual components.
NOTE 3 The tests only apply to ankle-disarticulation prostheses which include (foot) components of prosthetic anklefoot devices taken from the normal production line.
NOTE 4 The distal part comprises the knee unit, the ankle-foot device and all parts between. Tests on hip units are described in ISO 15032.