NACE SP0508
NACE SP0508 2010-AUG-19 Methods of Valdatng Equvalence to SO 8502-9 on Measurement of the Levels of Soluble Salts-tem No 21134
NACE SP0508 2010-AUG-19 Methods of Valdatng Equvalence to SO 8502-9 on Measurement of the Levels of Soluble Salts-tem No 21134
General
ISO 8502-9 Range of Variance
The assessment and determination of surface contamination (by salts) prior to application of protective coatings is critical to their service life expectancy. Determination of the level of surface cleanliness is performed using a field method for the conductometric determination of water-soluble salts in accordance with ISO 8502-9. As a part of that method, the extraction of soluble salt contaminants for analysis is performed in accordance with ISO 8502-6—the Bresle method. The field execution of this method involves using a syringe to inject deionized water into the Bresle patch, washing the substrate surface inside the patch, then extracting the test water for direct measurement of conductivity. Details of this procedure are provided in Appendix A (mandatory). Once a conductivity value is determined by the conductivity meter, expressed as microsiemens per centimeter (µS/cm), ISO 8502-9 provides a procedure to calculate the equivalent mass of the surface concentration as total surface density of the salts.
Step-by-step execution of the ISO 8502-6 and ISO 8502-9 test methods introduces a number of potential variances during the field evaluation. Some examples include background salt contamination in the Bresle patch itself, the sensitivity and resolution of the conductivity meter, cycle time of the test water inside the Bresle patch, human errors in measuring and injecting the test water, and general operator experience. These individual process variances combine to create an overall variance in the ISO 8502-9 test method.
The test method defined in ISO 8502-9 allows certain execution parameters of the test (e.g., size of adhesive patch, volume of test solution, and time of dissolution of salts inside the adhesive patch—the dwell time) to vary. This standard adopted best practices and has set the parameters in Appendix A. With a dwell time at a minimum of 90 seconds, the variance from operator error was found to be reduced dramatically when the operator timing is not perfect. Hence, a dwell time of 90 seconds was adopted in Appendix A. For the purpose of defining equivalence, the procedure in Appendix A shall be used to create the reference values to which the candidate method will be validated.
Bresle patches from different manufacturers or manufacturing lots may have different levels of background contamination. Specifications for salt contamination include the contribution of this background contamination so that the operator in the field does not have to subtract a background value to report the test results. Appendix A therefore includes a provision to correct the ISO 8502-9 test results to a consistent offset contributed by Bresle patch background contamination so that this standard gives consistent results when tests are performed with Bresle patches from different manufacturers or manufacturing lots.
The range of variance in ISO 8502-6 and 8502-9 has been demonstrated by extensive laboratory tests. The precision of a single ISO 8502-9 test result was determined to be ± 8.2 mg/m2 in the salt level range of 30 to 80 mg/m2. The absolute variance, and not the relative or percent variance, was found to be constant in this range. See Appendix B (nonmandatory) for details about these tests and the data analysis.
Any equivalent tools, methods, or procedures must show that they meet the same criteria, thereby providing the same measures of soluble salt level, and be within the same range of variance, as would be produced by following the procedure in ISO 8502-9.
The objective of this standard is to determine whether methods other than the Bresle method are suitable alternatives for measuring salt contamination in the field. Although tests may be performed on flat, horizontal surfaces for ease of use, each method should also be capable of performing measurements on vertical and overhead surfaces. When such measurements cannot be performed on vertical or overhead surfaces, this limitation must be noted in the validation report (Section 3).
Reasons for Salt Measurement
It is well known that applying protective coatings over salt contamination may cause degradation of the coating/substrate interface. This may lead to corrosion of the substrate and detachment of the coating. These effects reduce the expected service life of the coating.
Different salt ions influence the rate of corrosion under coatings differently. However, most common salts promote corrosion and have a negative effect on the service life of coatings.
Salts on a surface under a coating can create an osmotic cell, drawing water into and through the coating. The osmotic pressure generated can cause blisters to form in the coating.
The general relationship between the amount of salt contamination under the coating film and consequent deterioration of the performance may be considered to be linear.
Although virtually unobtainable, a zero measure of salt contamination may be considered the only "safe" level before application of coating films. Therefore, the maximum quantity of salt contamination allowed in a specification should be based on a holistic risk/cost analysis.